# Install the Test Bench

## Choose an appropriate location <a href="#choosing-an-appropriate-location" id="choosing-an-appropriate-location"></a>

### Structural requirements <a href="#structural-requirements" id="structural-requirements"></a>

The OP1400PR test benches are fairly **heavy sets of equipment** (up to **600 kg** for the heaviest models when incorporating optional equipment), thus care should be taken when choosing the installation location for the test bench. OPAL-RT highly recommends that the building structure where the bench will be sitting is designed to **support at least one metric tonne (1000 kg) of weight**.

### Space required <a href="#available-space" id="available-space"></a>

{% columns %}
{% column width="58.333333333333336%" %}
The final location for the OP1400PR test bench should consider the **dimensions of the cabinet**.

Please note that the **two side panels** and the **rear panel** are **hinged doors** that **may need to be open occasionally** to reconfigure wiring or for maintenance reasons. Although it is not necessary to provide permanent clearance for the side and rear doors, it is recommended to at least confirm that there is **enough space** and **sufficient wiring slack** to **roll the chassis**. This allows **access to the doors**.

The OP1400PR test bench cabinet is equipped with a **fan installed at the top** for ventilation. Therefore, **the top of the cabinet must not be obstructed in any way that will disrupt the airflow from the fan**. There must be **at least 15 cm (6 inches) of free space** between the **ceiling** and the **top of the cabinet**.
{% endcolumn %}

{% column width="41.666666666666664%" %}

<figure><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FaKws7HjA1ta3Oa8Bs3pT%2Fimage.png?alt=media&amp;token=2fd534f1-effb-4e31-b5a9-078bbe01566e" alt=""><figcaption></figcaption></figure>

{% endcolumn %}
{% endcolumns %}

### Environmental conditions <a href="#environmental-conditions" id="environmental-conditions"></a>

OPAL-RT requires that the following environmental conditions be met at the installation site of the OP1400PR test bench.

| Parameter             | Ideal                                                     | Mandatory                                                         |
| --------------------- | --------------------------------------------------------- | ----------------------------------------------------------------- |
| **Site location**     | Inside a temperature-controlled laboratory or equivalent. | Indoors                                                           |
| **Temperature**       | Regulated between 20 °C and 30 °C (68 °F and 86 °F)       | Between 5 °C and 40 °C                                            |
| **Site altitude**     | No more than 2000 m                                       | No more than 2000 m                                               |
| **Relative humidity** | Regulated to 50% RH or less                               | Maximum 80% RH up to 31 °C decreasing linearly to 50% RH at 40 °C |
| **Pollution**         | IEC Pollution Degree 2 (or better)                        | IEC Pollution Degree 2 (or better)                                |

### Availability of electrical connections <a href="#availability-of-electrical-connections" id="availability-of-electrical-connections"></a>

* Two dedicated single-phase power bars (100…240 Vac, 10 A).
* Power terminals (3P[^1] + PE[^2]) for connection to a 380/400/415/480 Vac three-phase supply.

Refer to the section [**Connect the three-phase mains to the PDU**](https://www.opal-rt.com/docs/op1400pr/get-started/connect-the-test-bench.md#connection-from-the-three-phase-mains-to-the-pdu) for the circuit dimensioning requirements.

Please note that it is theoretically possible to have a negative net power flow (energy fed back to the grid) on the 3-phase AC power line. This could be done using, for example, a PV[^3] system or a small wind turbine on a DER[^4] input of an OP1400PR test bench. The user should note that not all electricity distribution companies will allow customers to feed energy to the grid and should contact their local electricity distributor before attempting to do so.

In addition, please note that OP1400PR-10 and OP1400PR-20 models will consume steady-state unbalanced currents on their 3-phase AC input, which could lead to penalties imposed on the user by the local electricity distributor.

Contact your [**local OPAL-RT sales representative**](https://www.opal-rt.com/find-local-contact/) for more information if required.

## Unpack the test bench <a href="#unpacking" id="unpacking"></a>

### Uncrate the test bench <a href="#op1400pr-test-bench-crate" id="op1400pr-test-bench-crate"></a>

The test bench cabinet and included components are received in a crate. The following procedure must be executed to safely unpack the test bench: [**Uncrate an OPAL-RT Cabinet**](https://www.opal-rt.com/docs/op1400pr/get-started/install-the-test-bench/uncrate-an-opal-rt-cabinet.md).

### Locate the parts

Once the crate has been opened, locate the following items and ensure they are all present and undamaged.

#### Common parts

<table data-search="false"><thead><tr><th valign="top">Item</th><th valign="top">Description</th><th valign="top">Part number</th><th valign="top">Image</th><th valign="top">Quantity</th></tr></thead><tbody><tr><td valign="top"><strong>Fiber-optic SFP transceiver</strong></td><td valign="top">Fiber-optic transceiver module 4.25 GBd 850 nm 3.3 V for installation on OPAL-RT simulators to enable SFP communication with OP8110 units.</td><td valign="top">T00-0315</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FxyRcOrE9kdj3kQr5uHHE%2Fimage.png?alt=media&amp;token=bb747d87-2a42-4fc2-a2d3-5938cf83294a" alt="" data-size="original"></td><td valign="top">OP1400PR-10: 1<br>OP1400PR-20: 1<br>OP1400PR-30: 2<br>OP1400PR-60: 3</td></tr><tr><td valign="top"><strong>Fiber-optic cable</strong></td><td valign="top">Multimode duplex fiber-optic cable 62.5/125 LC-LC 2 meters (6.6 ft) to connect OPAL-RT simulator to OP8110 through SFP interface</td><td valign="top">T00-0797</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2F1yLgXg7B4jXFLNCUSaFF%2Fimage.png?alt=media&amp;token=d1028f3a-c6c8-4c73-b42f-a5c95e92fa18" alt="" data-size="original"></td><td valign="top">OP1400PR-10: 1<br>OP1400PR-20: 1<br>OP1400PR-30: 2<br>OP1400PR-60: 3</td></tr><tr><td valign="top"><strong>Ethernet cable</strong></td><td valign="top">CAT6 (500 MHz) UTP network cable - blue 10 ft (3 m)</td><td valign="top">T00-0817</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FEp9eN0BNaxBgDMN2mbr3%2Fimage.png?alt=media&amp;token=2b596f48-573b-48cb-97dd-9ff76ce78620" alt="" data-size="original"></td><td valign="top"><strong>For OP8110 connection  (pre-installed)</strong>: <br>OP1400PR-10: 1<br>OP1400PR-20: 1<br>OP1400PR-30: 2<br>OP1400PR-60: 3<br><br><strong>Additionally, for all test bench models</strong>: <br>1x Ethernet cable included within the crate for connection to the network switch inside the cabinet, &#x26;<br>1x Ethernet cable for simulator (if present).</td></tr><tr><td valign="top"><strong>Optical synchronization cable</strong></td><td valign="top">Plastic Optical Fiber (POF) patch cords 0.65 meter (25.5 in)</td><td valign="top">T00-6615</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2Fln1FhR4AR2nsklbjW6yZ%2Fimage.png?alt=media&amp;token=e15f7bbc-d4be-4b90-bf8c-93c5fa507feb" alt="" data-size="original"></td><td valign="top">OP1400PR-10: 0<br>OP1400PR-20: 0<br>OP1400PR-30: 1<br>OP1400PR-60: 2</td></tr><tr><td valign="top"><strong>Mini USB cable</strong></td><td valign="top">Cable USB A to right-angle Mini B, Black, 3 ft (91.5 cm)</td><td valign="top">T00-0805</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FyTBJUmUdAmZt0d8jxjZr%2Fimage.png?alt=media&amp;token=0dc79df4-3c0b-4204-8315-972695b50fe6" alt="" data-size="original"></td><td valign="top">1</td></tr><tr><td valign="top"><strong>Ring terminals to PowerLok (AC)</strong></td><td valign="top">Cable connecting the <a data-footnote-ref href="#user-content-fn-5">PDU</a> AC_OUT to the ITECH input power</td><td valign="top">113-1731</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FwX2fBF0EuPRMyVtvapKZ%2Fimage.png?alt=media&amp;token=d038990b-8915-41f5-b9ac-31b44f306822" alt="" data-size="original"></td><td valign="top">OP1400PR-10: 1<br>OP1400PR-20: 1<br>OP1400PR-30: 3<br>OP1400PR-60: 3<br><br><strong>Note</strong>: These come installed within the cabinet.</td></tr><tr><td valign="top"><strong>Auxiliary power cable</strong></td><td valign="top">Cable connecting from the PDU AUX to the amplifiers AUX</td><td valign="top">113-1732</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2Fxt8SNeEQzfBPAtV6moCH%2Fimage.png?alt=media&amp;token=6d004291-8f8a-475b-b49d-7dc56e8ed8da" alt="" data-size="original"></td><td valign="top">OP1400PR-10: 1<br>OP1400PR-20: 2<br>OP1400PR-30: 3<br>OP1400PR-60: 6<br><br><strong>Note</strong> - These come installed within the cabinet.</td></tr><tr><td valign="top"><strong>25 cm red banana cable</strong></td><td valign="top">Banana cable connecting the amplifier DC+ inputs of group 1 and group 2</td><td valign="top">T00-9659</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FkGocpplFhMC1xPWrAgpL%2Fimage.png?alt=media&amp;token=6eb90c33-921a-455e-9fb7-725d33f64811" alt="" data-size="original"></td><td valign="top">OP1400PR-10: 0<br>OP1400PR-20: 1<br>OP1400PR-30: 0<br>OP1400PR-60: 3</td></tr><tr><td valign="top"><strong>25 cm black banana cable</strong></td><td valign="top">Banana cable connecting the amplifier DC- inputs of group 1 and group 2</td><td valign="top">T00-9658</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2F0mHCCIn6U1CseCOZzuGX%2Fimage.png?alt=media&amp;token=eee8a4b5-42c9-4978-aff6-a082645b97cb" alt="" data-size="original"></td><td valign="top">OP1400PR-10: 0<br>OP1400PR-20: 1<br>OP1400PR-30: 0<br>OP1400PR-60: 3</td></tr><tr><td valign="top"><strong>25 cm white banana cable</strong></td><td valign="top">SKL425-E/N 4 mm safety test lead 25 cm (9.8 in), white</td><td valign="top">T00-7535<br>(or T00-8896)</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FajjByNoMxiicL2DNk389%2Fimage.png?alt=media&amp;token=7baddbf2-77b4-4470-a0b7-eac951122a87" alt="" data-size="original"></td><td valign="top">OP1400PR-10: 3<br>OP1400PR-20: 6<br>OP1400PR-30: 9<br>OP1400PR-60: 18</td></tr><tr><td valign="top"><strong>2 meters banana cable harness</strong></td><td valign="top">SLK425-E/N 4 mm safety test lead, 2 meters (6.5 ft), 4 cable harnesses (red, black, blue and white)</td><td valign="top">113-1371</td><td valign="top"><p></p><p><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FmRtvFLkVi7VksOG3LtIc%2Fimage.png?alt=media&amp;token=5d1af6b6-304e-41fc-be92-808d2d905b38" alt="" data-size="original"></p></td><td valign="top">OP1400PR-10: 2<br>OP1400PR-20: 4<br>OP1400PR-30: 6<br>OP1400PR-60: 12</td></tr><tr><td valign="top"><strong>VFEM cables</strong></td><td valign="top">Remote measurement cables for the OP8110.</td><td valign="top">113-1427</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2Ft9NKiFkC2oecR7KG1rqn%2Fimage.png?alt=media&amp;token=fd21b569-d2fb-4c27-94c3-71b834121c6f" alt="" data-size="original"></td><td valign="top">OP1400PR-10: 1 set<br>OP1400PR-20: 2 sets<br>OP1400PR-30: 3 sets<br>OP1400PR-60: 6 sets</td></tr><tr><td valign="top"><strong>ITECH Synch cable</strong></td><td valign="top">On/Off synchronization cable</td><td valign="top">113-1588</td><td valign="top"><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2F0i4di0s5wqeqvP4M1NTs%2Fimage.png?alt=media&amp;token=e74b6d78-2833-4cc4-b3e3-05848fad43eb" alt="" data-size="original"></td><td valign="top"><p>These cables are shipped with the ITECH connectors attached.</p><p>The number of cables differs depending on the bench configuration:</p><p>OP1400PR-10: 0<br>OP1400PR-20: 0<br>OP1400PR-30: 2<br>OP1400PR-60: 2</p></td></tr></tbody></table>

#### Additional parts

Additional components for test benches that have either or both the **LP  bus bar box** and the **HP bus bar box**.

{% hint style="info" %}
**-LP** configuration refers to any OP1400PR PHIL Prime test bench that has an LP bus bar box installed. For example, OP1400PR-10-LP, OP1400PR-20-LP and so on.

**-HP** configuration refers to any OP1400PR PHIL Prime test bench that has an HP bus bar box installed. For example, OP1400PR-30-HP and OP1400PR-60-HP.

**-LPHP** configuration refers to any OP1400PR PHIL Prime test bench that has both an LP bus bar Box AND an HP bus bar box installed. For example OP1400PR-30-LPHP and OP1400PR-60-LPHP.
{% endhint %}

| Item                                                              | Description                                                                                                                                    | Part number                     | Image                                                                                                                                                                                                                                                     | Quantity                                                                                                            |
| ----------------------------------------------------------------- | ---------------------------------------------------------------------------------------------------------------------------------------------- | ------------------------------- | --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | ------------------------------------------------------------------------------------------------------------------- |
| <p><strong>VFEM cable</strong><br><strong>LP bur bar</strong></p> | VFEM sensor, banana connector, 80 in (203 cm), used with the LP bus bar                                                                        | 113-1428                        | <img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FHAYOfizmaU6pRjFdCMaG%2Fimage.png?alt=media&amp;token=02e044c4-8056-4825-ac25-0d0625a09c28" alt="" data-size="original"> | <p><strong>-LP</strong> config : 2 sets<br><strong>-LPHP</strong> config : 2 sets<br><br></p>                       |
| **VFEM cable**                                                    | VFEM sensor,  ring connector, 120 in (305 cm), used with the HP bus bar                                                                        | 113-1429                        | <img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2Ff4gXqQIGxYreWtq0gNYx%2Fimage.png?alt=media&amp;token=f28cb09a-cccf-4679-a2f2-a06f8ac1a548" alt="" data-size="original"> | <p><strong>-HP</strong> config: 2 sets<br><strong>-LPHP</strong> config: 2 sets</p>                                 |
| **50 cm single banana cable red**                                 | SKL425-E/N, 4 mm safety test lead, 50 cm (20.7 in), red                                                                                        | T00-7239                        | <img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2Fzxf9eFtX0wDzMSZvCmWP%2Fimage.png?alt=media&amp;token=2e19bf88-3793-4ba9-ae73-76cfe4d57fb6" alt="" data-size="original"> | <p><strong>-LP</strong> config: 4<br><strong>-HP</strong> config: 4<br><strong>-LPHP</strong> config: 4<br><br></p> |
| **50 cm single banana cable black**                               | SKL425-E/N, 4 mm safety test lead,  50 cm (20.7 in), black                                                                                     | T00-7240                        | <img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FrwaHz93S8X8GNycp11Un%2Fimage.png?alt=media&amp;token=0f7bb0f8-1d5c-4c00-bca4-134d25eab800" alt="" data-size="original"> | <p><strong>-LP</strong> config: 4<br><strong>-HP</strong> config: 4<br><strong>-LPHP</strong> config: 4</p>         |
| **ITECH fiber-optic cable**                                       | ITECH fiber-optic cable for IT6000 and IT8000                                                                                                  | <p>T00-8856</p><p>(IT-E168)</p> | <img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FqIK7v5FVEyutcQl69wAx%2Fimage.png?alt=media&amp;token=5e023e56-8f3e-476a-8ff8-be37b97cc8ec" alt="" data-size="original"> | <p><strong>-LP</strong> config: 3<br><strong>-HP</strong> config: 3<br><strong>-LPHP</strong> config: 3</p>         |
| **Bus bar control power 24VDC**                                   | <p>Brings 24 VDC power from the E-Stop Module to the bus bar box.</p><p></p><p>Multiple bus bar boxes will be connected in daisy chaining.</p> | 113-1302                        | <img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2Fa64YScwhFemAB8lc43zb%2Fimage.png?alt=media&amp;token=f46d2d56-ae3c-4a20-9f3d-f4bb90c08eed" alt="" data-size="original"> | <p><strong>-LP</strong> config: 1                                                                                   |
| <br><strong>-HP</strong> config: 1                                |                                                                                                                                                |                                 |                                                                                                                                                                                                                                                           |                                                                                                                     |
| <br><strong>-LPHP</strong> config: 2</p>                          |                                                                                                                                                |                                 |                                                                                                                                                                                                                                                           |                                                                                                                     |

#### Separately packaged equipment <a href="#separately-packaged-equipment" id="separately-packaged-equipment"></a>

To minimize the risk associated with shipping and handling, the following components are removed from the test bench before shipping and are packaged separately:

* [OP1400PR PHIL Prime - OP8110 4Q power amplifier](https://www.opal-rt.com/docs/op1400pr/op1400pr-overview/description-and-components/op8110-4q-power-amplifier.md)
* [OP1400PR PHIL Prime - DC power supplies ITECH IT6000C Series](https://www.opal-rt.com/docs/op1400pr/op1400pr-overview/description-and-components/dc-power-supply.md)

The following quantities of each item should be expected depending on the purchased test bench model:

|                    | OP1400PR-10 models | OP1400PR-20 models | OP1400PR-30 models | OP1400PR-60 models |
| ------------------ | :----------------: | :----------------: | :----------------: | :----------------: |
| **OP8110-3**       |          1         |          -         |          1         |          -         |
| **OP8110-6**       |          -         |          1         |          1         |          3         |
| **IT6006C-500-30** |          1         |          -         |          3         |          -         |
| **IT6012C-500-60** |          -         |          1         |          -         |          3         |

Ensure that **all these components** have been **received**.

Once unpacked, ensure that **all items** are **undamaged**.

[^1]: Three phases

[^2]: Protective Earth

[^3]: Photovoltaic

[^4]: Distributed Energy Resource

[^5]: Power Distribution Unit


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://www.opal-rt.com/docs/op1400pr/get-started/install-the-test-bench.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.