> For the complete documentation index, see [llms.txt](https://www.opal-rt.com/docs/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://www.opal-rt.com/docs/op1400pr/op1400pr-overview/description-and-components/emergency-stop-module.md).

# Emergency-stop Module

<figure><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2F97TyhmDSaOIABPOpFvcd%2Fimage.png?alt=media&amp;token=a925abf4-dda3-4dec-9e45-3b1bece1233e" alt=""><figcaption><p>Front view of the OP1400PR E-stop module</p></figcaption></figure>

## Description <a href="#description" id="description"></a>

The emergency-stop, or E-stop,  module provides the following features to the OP1400PR PHIL Prime Series test benches:

* **Start/Stop button** to activate or deactivate the 3-phase 380 V/480 V supply to the DC power supplies.
* **Status indication LED** for the **24 V PDU** (Power Distribution Unit) within the **PDU** (Power Distribution Unit).
* **Emergency stop push button** to quickly isolate the test bench's power path (excluding single-phase power bars) from all energy sources.

## Controls and indicators <a href="#controls-and-indicators" id="controls-and-indicators"></a>

The controls and indicators are all located at the front of the module.

<figure><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FdJD0zA9eZvz7DnddYKk6%2Fimage.png?alt=media&amp;token=9d03e72f-8d1a-4043-9b62-8b366531eb1f" alt=""><figcaption><p>Start/Stop buttons, 24 V ON LED, and emergency-stop button</p></figcaption></figure>

### Start/Stop buttons <a href="#start-stop-button" id="start-stop-button"></a>

These buttons allow you to control the 3-phase 400 V/480 V supply to the DC power supplies of the test bench. An indicator light located between the Start and Stop buttons lights up when activated.

### 24 V ON LED <a href="#id-24v-on-led" id="id-24v-on-led"></a>

This LED indicates the status of the 24 V DC power supply within the **PDU**. Note that this LED will remain ON even if the emergency-stop button has been pressed.

### Emergency-stop button <a href="#emergency-stop-button" id="emergency-stop-button"></a>

The emergency-stop button allows you to quickly secure the test bench in **emergencies only**.

{% hint style="danger" %}
It should **never be used as a power-off button**.
{% endhint %}

It is **hard-wired to an emergency-stop loop that energizes the contactors** inside the PDU and bus bar modules (if any) in the test bench. The loop keeps contactors energized during normal operation, then de-energizes them when activated.

When the user presses the emergency-stop button, all these contactors open effectively isolating the test bench's power path (excluding single-phase power bars) from all possible sources of electrical energy.

The button remains depressed after activation. Rotate it clockwise to reset the emergency-stop loop. The emergency-stop button can be secured with a padlock if desired.

## Connectors

The connectors are all located at the back of the module.

<figure><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2FDw7xuTozWlwlHHJOp1i7%2Fimage.png?alt=media&amp;token=d852701d-2976-4de5-b7da-751324dab19a" alt=""><figcaption><p>Rear P1 and P2 connectors</p></figcaption></figure>

### P1 connector <a href="#p1-connector" id="p1-connector"></a>

The P1 connector interfaces with the PDU’s P1 connector via the cable 113-1304 - Cable E-stop to PDU.

### P2 connector <a href="#p2-connector" id="p2-connector"></a>

The P2 connector interfaces with either the LP[^1] or HP[^2] BUS BAR via cable 113-1302 - Cable aux PS to bus bar 24 V DC E-stop.

## Electrical schematic <a href="#electrical-schematic" id="electrical-schematic"></a>

<figure><img src="https://3155739458-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FzVJdfp93TWLISP6nkqTs%2Fuploads%2Fgv3Pekc2qS64BT2VfosP%2Fimage.png?alt=media&amp;token=89f9ca50-d71b-4480-a7a4-0db45fa132c7" alt=""><figcaption><p>E-stop electrical schematic</p></figcaption></figure>

&#x20;

[^1]: Low power

[^2]: High power


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://www.opal-rt.com/docs/op1400pr/op1400pr-overview/description-and-components/emergency-stop-module.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.